LETTER TO EDITOR
Year : 2013  |  Volume : 5  |  Issue : 2  |  Page : 143-144  

Effect of different anticoagulants on HBA1C estimation and its stability


Department of Biochemistry, R D Gardi Medical College, Ujjain, Madhya Pradesh, India

Date of Web Publication15-Oct-2013

Correspondence Address:
Booloo Sharma
Department of Biochemistry, R D Gardi Medical College, Ujjain, Madhya Pradesh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0974-2727.119871

Rights and Permissions

How to cite this article:
Sharma B, Sarmah D, Sonker P. Effect of different anticoagulants on HBA1C estimation and its stability. J Lab Physicians 2013;5:143-4

How to cite this URL:
Sharma B, Sarmah D, Sonker P. Effect of different anticoagulants on HBA1C estimation and its stability. J Lab Physicians [serial online] 2013 [cited 2019 Dec 7];5:143-4. Available from: http://www.jlponline.org/text.asp?2013/5/2/143/119871

Sir,

Glycated hemoglobin/HbA1c is always considered as a stable indicator of glycaemia for the preceding three months. [1] Its potential utility in diabetic care was first reported in 1985 WHO report, and by 2010 all the major expert committee and association across the globe including the ADA has recommended HbA1c for the diagnosis of Type 2 DM, besides its role in prognosis. [2] So, the importance of HbA1c estimation in diabetes has increased manifold in recent years.

Most of the commercial kits for HbA1c estimation requires sample to be collected in EDTA anticoagulant. This requires additional sample collection for the patient. So, we estimated HbA1c in sample collected in K 3 EDTA, Na-citrate, Na-heparin and Na-fluoride/Na 2 EDTA vials separately and then to see for any variation. All the vials used are of BD (Becton, Dickinson and Company) Vacutainer manufactured by Becton, Dickinson India. HbA1c is estimated after 2-3 hours of blood collection by cation exchange HPLC based BioRad D 10 analyzer. It is said that HbA1c has high pre-analytical stability and is stable for one week when stored at 4°C and for one year when stored at −70°C. [3],[4] So, to verify the stability of HbA1c in the above vials HbA1c is estimated for 7 days in the same vial. Just after estimation, vials are stored at 4 o C and then HbA1c measured every day till 7 th day in all the vials, marking 1 st day as 0 day.

As indicated by [Figure 1],[Figure 2],[Figure 3],[Figure 4],[Figure 5] and [Figure 6], there is no significant change in the HbA1c values estimated in the vials containing different anticoagulant. This is in accordance with the findings of Mailankot et al. [5] [Figure 1],[Figure 2],[Figure 3],[Figure 4],[Figure 5] and [Figure 6], also shows that the stability of HbA1c is not altered in the vials containing different anticoagulant when stored at 4°C for 7 days. The findings withdraws the need for a separate EDTA sample for HbA1c estimation as advocated by various commercial kits, and it also ensures the stability of the collected sample for 7 days when stored at 4°C.
Figure 1: The results for healthy non-diabetic adult samples

Click here to view
Figure 2: The results for healthy non-diabetic adult samples

Click here to view
Figure 3: The results for healthy non-diabetic adult samples

Click here to view
Figure 4: The results for adult diabetic samples

Click here to view
Figure 5: The results for adult diabetic samples

Click here to view
Figure 6: The results for adult diabetic samples

Click here to view


To conclude, we can infer that the unnecessary extra sample collection for estimation of HbA1c could be avoided which in turn can improve patient compliance and this may contribute towards a better diabetic care. It can also be accounted that the avoidance of an extra Vacutainer for HbA1c can reduce the cost of HbA1c test, which is undoubtedly a costly test.

 
   References Top

1.Fitzgibbons JF, Koler RD and Jones RT, ibid, 1976;58:820-4.  Back to cited text no. 1
    
2.American Diabetes Association. Diagnosis and classification of diabetes mellitus. Diabetes Care 2012;35(Suppl 1):S64-71.  Back to cited text no. 2
    
3.Sacks DB, Bruns DE, Goldstein DE, Maclaren NK, McDonald JM, Parrott M. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2002;48:436-72.  Back to cited text no. 3
    
4.Little RR, Rohlfing CL, Tennill AL, Connolly S, Hanson S. Effects of sample storage conditions on glycated hemoglobin measurement: Evaluation of five different high performance liquid chromatography methods. Diabetes Technol Ther 2007;9:36-42.  Back to cited text no. 4
    
5.Mailankot M, Thomas T, Praveena P, Jacob J, Benjamin JR, Vasudevan DM et al. Various anticoagulants and fluoride do not affect HbA1C level. Ind J Clin Biochem 2012;27:209.  Back to cited text no. 5
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]



 